CALIBRATION LABORATORY OF CLRP
Nowadays, the quality management system ISO 9001:2015 for the industrial and manufacturing and the laboratory quality management system ISO/IEC 17025:2017 for the laboratory able to ensure the consistency of product produced and the effectiveness of continuous improvement. One of the requirements of ISO 9001:2015 and ISO 17025:2017 states that organizations need to calibrate equipments that affect the measurement results as evidence of measurement traceability.
In order to improve the ability of competition in the global market, Center for Leather, Rubber, and Plastics (CLRP) has a Calibration Laboratory accredited by the National Accreditation Body of Indonesia (KAN) with id number LK-085-IDN. The Calibration Laboratory of CLRP oriented to customer satisfaction in helping the growth of the business community, especially small, medium, and large industry.
To discharge its activities, the Calibration Laboratory of CLRP implemented the laboratory quality management system ISO/IEC 17025:2017. National Accreditation Body of Indonesia (KAN) provides some scope that have been accredited, namely: mass, temperature, volumetric apparatus, force, pressure, instrumentation analyzer, dimensions, and time & frequency.
Our Motto: Professional, Reliable, and Excellent in Service
SCOPE OF CLRP CALIBRATION LABORATORY
1. MASSA
NO | TOOLS | METHOD | CAPABILITY | NOTES | COST (IDR) | COST NOTES |
---|---|---|---|---|---|---|
1 | Mechanic / electronic balances | NMI Australia, Monograph 4: The Calibration of Weights and Balances (May 2010) | up to 200 gram | Resolution until 0,1 mg | 300.000 | |
up to 2000 gram | Resolution until 10 mg | 300.000 | ||||
up to 200 kg | 350.000 | For 10 ~ 150 Kg | ||||
500.000 | For > 150 Kg | |||||
2 | Scales | OIML R 111-1 (2004) | 0,001 gram ~ 2000 gram | Scales class M | 120.000 | per pcs |
2. TEMPERATURE
NO | TOOLS | METHOD | CAPABILITY | NOTES | COST (IDR) | COST NOTES |
---|---|---|---|---|---|---|
1 | Liquid-in-glass thermometer | SNSU PK.S-01:2020 | 0 ~ 100 °C | Resolution until 0,4 °C | 250.000 | |
2 | Dial/digital thermometer | SNSU PK.S-02:2021 | 0 ~ 100 °C | Resolution until 0,1 °C | 250.000 | Dial |
450.000 | Digital | |||||
3 | COD reactor / Thermoreactor | MK. IIId.S.LPK No. 7.4 Revision 2/2 (09 September 2021) | 50 ~ 200 °C | Resolution until 0,1 °C | 430.000 | |
4 | Hotplate / Rheometer / Mooney Viscometer / Plastimeter | MK. IIId.S.LPK No. 7.5 Revisionon 2/2 (3 October 2022) | 50 ~ 250 °C | Resolution until 0,1 °C | 350.000 | |
5 | Freezer | KAN Pd-02.04 Revisionon: 0 (2019) | -25 ~ 0 °C | Resolution until 0,1 °C | 500.000 | |
6 | Refrigerator / Showcase / Sample cabinet | KAN Pd-02.04 Revision: 0 (2019) | -5 ~ 20 °C | Resolution until 0,1 °C | 500.000 | |
7 | Incubator | KAN Pd-02.04 Revision: 0 (2019) | 0 ~ 100 °C | Resolution until 0,1 °C | 500.000 | |
8 | Waterbath | KAN Pd-02.04 Revision: 0 (2019) | 0 ~ 100 °C | Resolution until 0,1 °C | 600.000 | |
9 | Oven | KAN Pd-02.04 Revision: 0 (2019) | 30 ~ 200 °C | Resolution until 0,1 °C | 500.000 | |
10 | Muffle Furnace | KAN Pd-02.04 Revision: 0 (2019) (Modification) | 30 ~ 700 °C | Resolution until 0,1 °C | 650.000 |
3. VOLUMETRIC APPARATUS
NO | TOOLS | METHOD | CAPABILITY | NOTES | COST (IDR) | COST NOTES |
---|---|---|---|---|---|---|
1 | Buret | ASTM E 542 - 01 (2021) | 10 ml ~ 100 ml | 184.000 | ||
2 | Measuring glass | 10 ml ~ 1000 ml | 177.000 | For ≥ 100 ml | ||
156.000 | For < 100 ml | |||||
3 | Measuring flask | 5 ml ~ 1000 ml | 150.000 | |||
4 | Measuring pipette | 0,1 ml ~ 25 ml | 185.000 | |||
5 | Pycnometer | 5 ml ~ 100 ml | 150.000 | |||
6 | Volumetric Pipette | 1 ml ~ 100 ml | 167.000 |
4. FORCE
NO | TOOLS | METHOD | CAPABILITY | NOTES | COST (IDR) | COST NOTES |
---|---|---|---|---|---|---|
1 | Tensile strength / Universal testing machine | SNSU PK.M-03:2021 | 0 ~ 500 kgf | 500.000 |
5. PRESSURE
NO | TOOLS | METHOD | CAPABILITY | NOTES | COST (IDR) | COST NOTES |
---|---|---|---|---|---|---|
1 | Pressure gauge | DKD-R 6-1 edition 03/2014; Euramet CG-17 version 4.0 (04/2019) | 0 ~ 5000 Psi | 300.000 |
6. INSTRUMENTATION ANALYZER
NO | TOOLS | METHOD | CAPABILITY | NOTES | COST (IDR) | COST NOTES |
---|---|---|---|---|---|---|
1 | pH Meter | ASTM E70-19 | 4, 7, 10 | 255.000 | ||
2 | UV-VIS Spectrophotometer | SNSU PK.F-01:2020 | Wavelength accuracy (Holmium) : 279~638 nm | 500.000 | ||
Wavelength accuracy (Didynium) : 443~887 nm | ||||||
Photometric Linearity at α 590 nm : 0,2 Abs; 0,5 Abs; 1,0 Abs |
7. DIMENSIONS
NO | TOOLS | METHOD | CAPABILITY | NOTES | COST (IDR) | COST NOTES |
---|---|---|---|---|---|---|
1 | Caliper | JIS B 7507-2016 | 0 ~ 225 mm | Manual, dial, digital | 234.000 | |
2 | Outside micrometer | JIS B 7502-2016 | 0 ~ 50 mm | Manual, dial, digital | 500.000 | |
3 | Thickness gauge | ASME B89.1.10M-2001 | 0 ~ 30 mm | Manual, dial, digital | 250.000 | |
4 | Depth Gauge | JIS B 7518:2018 | 0 ~ 100 mm | Manual, dial, digital | 400.000 |
8. TIME AND FREQUENCY
NO | TOOLS | METHOD | CAPABILITY | NOTES | COST (IDR) | COST NOTES |
---|---|---|---|---|---|---|
1 | Stopwatch | NIST SP 960-12 (2009) | 10 seconds ~ 32400 seconds | 150.000 | ||
2 | Timer | NIST SP 960-12 (2009) | 15 seconds ~ 7200 seconds | 150.000 | ||
3 | Centrifuge / Stirer | MK.IIId.WF.LPK No. Part 7 (Revision: 1/2-22 July 2022) | 50 rpm ~ 5000 rpm | 280.000 | ||
4 | Speedometer | MK.IIId.WF.LPK No. Part 8 (Revision: 1/2-22 July 2022) | 5 mm/min ~ 500 mm/min | 280.000 | ||
20 km/hour ~ 300 km/hour |
Scope that has not been accredited by KAN
NO | TOOLS | METHOD | CAPABILITY | NOTES | COST (IDR) | COST NOTES |
---|---|---|---|---|---|---|
1 | Turbidimeter | STABLCAL-Instructions for Use (HACH, Ed. 5, October 2009) | 0,05 NTU; 15,1 NTU; 100 NTU; 751 NTU | 300.000 | ||
2 | TDS meter | OIML R 68 (1985) | 7,5 mg/L; 74,6 mg/L; 746 mg/L | 250.000 | ||
3 | Conductivity meter | OIML R 68 (1985) | 0,015 mS/cm; 0,14 mS/cm; 1,41 mS/cm | 250.000 | ||
4 | Autoclave | BS 2646-1:2021 | Suhu & Tekanan | 600.000 |